Control of cleaning solutions by surface tension measurements
Background
Compared with organic solvents, cleaning with aqueous surfactant solutions has less impact on objects and is safer for workers. Hence, it is widely used in households and industrial fields.
Even with a small amount of surfactant, water's surface tension decreases significantly, increasing its permeability while enveloping and drawing oil away. However, the surface tension remains constant after a particular concentration (critical micelle concentration, aka CMC), and the detergency stabilizes. It is crucial to manage the detergency quantitatively to always use a cleaning solution in the best condition.
Problems to be solved
The surfactant of cleaning and plating solution is consumed over time, decreasing detergency and permeation. It is essential to periodically replenish the tank with a surfactant or replace the liquid. Since increasing the amount of the surfactant added does not improve the detergency, the amount used must be properly controlled. Therefore, we tested the possibility of controlling the concentration by measuring surface tension.
Measurements and results
We conducted the test using our surface tensiometer, model DY-700, with the optional automatic CMC measurement kit. We measured using the Wilhelmy method, in which a platinum plate is brought into contact with a liquid sample placed in a petri dish. The surface tension is calculated from the pulling force toward the liquid, detected by a weighing system (Fig.1).
Figure 1 – Illustration of the surface tension measurement
Firstly, a value of CMC was evaluated by plotting a graph of the surfactant’s concentration in the cleaning solution and its surface tension (Fig.2). It can be seen that the value of the surface tension tends to increase as the concentration decreases with surfactant consumption in the cleaning solution. By controlling the surface tension value of the cleaning solution around the CMC, it is possible to maintain the detergency and set the amount of surfactant used.
Conclusion
Our surface tensiometer series, the DY-500 and DY-700, mounted with an automatic CMC measurement kit, is suitable for evaluating detergency by surface tension measurements to ensure the economical use of surfactants, the best reliability, and product quality.