1. 界面科学測器の協和界面科学
  2. Kyowa Interface Science
  3. Products
  4. Contact Angle Meter
  5. MCA-4

MCA-4

MCA-4: main instrument incl. microinjector system, control boxes, and observation monitor

MCA-3

MCA-4: sample stage and 5µm glass capillary

MCA-3

MCA-4: contact angle measurement of water on a human hair

MCA-3

MCA-4: sample stage with optional ink jet head system

MCA-3

MCA-4: top view observation of a water drop on a printed circuit board

MCA-3
MCA-3
MCA-3
MCA-3
MCA-3

Top-view camera

This video shows how the top-view camera helps to deposition the droplets with a glass capillary with 5µm inner diameter with pinpoint accuracy onto measuring points of a circuit inside a ROM and of a read-write head of a hard disk.

The two videos below show that besides the pinpoint deposition of the droplets the top-view camera can also be used for the observation of the three-phase contact line of the droplets. Here, droplets are deposited inside 100µm wide cells of an OLED panel. As the height of the cells' walls does not allow direct measurements of the contact angle from the side, the top-view camera helps to evaluate the surface cleanliness by observing the spreading behavior of the droplets on the solid surface.

  • Fast image capturing system with standard 400 fps at 640x480 resolution USB3.0 CMOS camera, max. 2700fps at 640x64 resolution
  • Automatic recognition of drop deposition by FAMAS analysis software
  • Top view camera allows for deposition of droplets with pinpoint accuracy and for observation of the spreading of micro-droplets
  • The high magnification lens and a 12-fold zoom provide a field of view from approximately 75x56µm to 910x680µm
  • Fine adjustable micro stage in x-, y- and z-axis for exact positioning
  • Fast image capture system allows accurate measurements of even quick evaporating pico-liter droplets
  • Droplet calibration standard bearing three sets of one full circle for calibration and three droplet silhouettes of 5º, 60º, and 108º for periodic inspection of measurement accuracy

Hardware

  • Software-controlled micro-stage
  • Jacket-type sample stage for temperature measurements from +10 to +70℃
  • Inkjet head system, using piezoelectric technology to generate droplets from about 8 to 60pL
  • Miniaturized areas such as of silicon wafer, OLED panels, electronic components, printed circuit boards, etc.
  • Very thin wire materials such as single fibers, hairs, filaments, etc.
  • Defined locations on the surface of tiny samples
  • Ink at very small droplet volume (emitted from an ink-jet head)
  • Other or micro-structured areas where conventional methods and droplet volume are not applicable
  MCA-4
Camera system




USB3.0 CMOS camera with:
400fps at 640x480 pixels
500fps at 640x400 pixels
1000fps at 640x200 pixels
2700fps at 640x64 pixels
Optical system


focus: manual focus with a 12-fold zoom lens
field of view: 75x56μm to 910x680μm ±5%
magnification: 5.3 to 64 times
Measuring methods contact angle: sessile drop (static and dynamic)
Analysis methods


contact angle: θ/2, tangent, curve fitting (ellipse, circle)
surface free energy of solids: OWRK, Owens-Wendt, Kaelble-Uy, Kitazaki-Hata, Wu, acid-base, Zisman
Measuring range contact angle: 0.1 to 180°
Resolution contact angle: 0.1°
Accuracy1) contact angle: 0.5°
Maximum sample surface area and height (WxDxH) 50x25x5mm

Sample stage operation:
X-axis
Y-axis
Z-axis
manual
25mm
25mm
15mm
Dispenser operation:
X-axis
Y-axis
Z-axis
Droplet generation
Air supply pressure
three-dimensional manual manipulator
30mm
30mm
30mm + 10mm hydraulic micro-manipulator
automatic
0.4 to 0.7 Mpa
Dimensions (WxDxH) of:
Main instrument
Rack for control boxes
Installation area

550x620x700mm
420x320x323mm
1000x650x720mm
Approximately weight of:
Main instrument
Control box

50.0kg
15.0kg
Electric power:
Voltage
Frequency
Power consumption

AC100-240V
50/60 Hz
60W 125VA
Operating environment

temperature: +10 to +35℃, humidity: 30-80%RH (non-condensing)
positioned away from sources of electrical noise and vibration

1) Accuracy is the repeatability in terms of standard deviation based on the manufacturer’s calibration standard